Low Flow Rotameter Flow Meter
For lab, dosing, and sample-injection service
DN15 and DN20 metal-tube variable-area
Water 0.025 to 2.5 m³/h or air 0.4 to 65 Nm³/h
Local pointer (no power, passive) or 4-20 mA
- Magnetic float in tapered metal tube; vertical mount, flow bottom-up, ±1.5 % FS standard
- Four DN15 float classes plus two DN20 classes give 1:10 turndown each on water service
- Pointer build is passive Ex-by-design; no electronics, no Ex certificate burden on classified areas

Small-bore metal-tube variable-area meter for lab benchtop water, chemical-injection-skid dosing, reactor sample lines, additive metering, and pilot-scale R&D service. DN15 and DN20 only. Water 0.025 to 2.5 m³/h across six float classes; air 0.4 to 65 Nm³/h across six classes. The pointer build is passive (no power, no Ex cert); 4-20 mA / HART is optional. For DN25+ bores go to the metal-tube-rotameter parent.
Benefits
- Small-bore DN15 / DN20 focus: six float classes across the two bores; sized for lab, dosing, and sample-injection without DN25+ overhead.
- ±1.5 % FS standard accuracy: factory-calibrated to water (20 °C) or air (20 °C, 101.325 kPa); ±2.5 % FS economy when budget rules.
- Passive pointer build is Ex-by-design: no electronics, no Ex certificate burden; the lowest-cost option for classified dosing skids.
- 304 / 316L / PTFE-lined / Hastelloy wetted parts: match material to clean water, light chemical, or aggressive acid / alkali dosing.
- Metric thread M20×1.5 / M36×1.5 small-bore: standard for DN15 / DN20; flange PN16 or PN25 by order if skid uses flanged manifolds.
- 4-20 mA / HART optional: electrical build with LCD, IP65, Ex d IIC T6 Gb when remote signal or DCS-tie is needed.
- Lead time 5–7 business days from the factory.
Configure your build.
Select your specs and the configurator returns the model code plus a ballpark unit price. Submit for engineering review and verified sizing within 4 business hours.
Your Configuration
✓ Medium and chemical composition reviewed by engineer
✓ Reply within 4 business hours
Typical applications
Four small-bore lab and dosing applications where the metal-tube low-flow build is the working pick.
Chemical-injection skid dosing trim
Per-line dosing indication on additive / inhibitor / scale-inhibitor skids. Pointer build gives direct visual confirmation on plant-walk.
Reactor sample-line indication
Continuous sample draw to analyser or lab. DN15 small-bore catches sample flow where DN25+ rotameter would be over-spec.
Pilot-scale and R&D rig
Bench-top pilot reactor, R&D loop, kinetic-study rig. Multi-float order swaps the float to match the campaign flow range.
Lab benchtop water service
Cooling-water indication on lab apparatus, jacketed reactor, condenser loop. Local pointer is the lowest-cost, no-power working pick.
Specifications
| Measurement principle | Variable-area; magnetic float rises in a vertically-mounted tapered metal tube under upward flow; external magnetic indicator reads float position |
|---|---|
| Pipe size range | DN15 (1/2″) and DN20 (3/4″) small-bore (DN25+ on parent metal-tube page) |
| Liquid flow range (water) | DN15: 0.025–0.25 / 0.04–0.4 / 0.063–0.63 / 0.1–1.0 m³/h. DN20: 0.16–1.6 / 0.25–2.5 m³/h |
| Gas flow range (air) | DN15: 0.4–7 / 0.7–10 / 1.1–16 / 2–25 Nm³/h. DN20: 3.5–40 / 5–65 Nm³/h |
| Reference standard | JB/T 9248-2015 (metal-tube variable-area flow meter) |
| Accuracy | ±1.5 % FS (code A) standard; ±2.5 % FS (code B) economy |
| Repeatability | ±0.5 % FS |
| Turndown | 1:10 per float (multi-float order delivers wider effective range by float swap at re-calibration) |
| Working pressure | Up to 2.5 MPa (DN15 / DN20 range) |
| Medium temperature | −40 to +120 °C standard; PTFE-lined build −40 to +85 °C; high-T option to +220 °C; jacketed option to +300 °C |
| Ambient temperature | −40 to +65 °C |
| Wetted parts | 304 stainless steel (standard) / 316L stainless steel / PTFE-lined / Hastelloy C / Titanium |
| Housing | Cast aluminium with stainless-steel measuring tube and external magnetic indicator housing |
| Indicator options | Local mechanical pointer (passive, no power, no Ex cert needed) / 4-20 mA transmitter / HART / RS-485 Modbus / Hi-Lo alarm contacts |
| Display | Mechanical scale on pointer build; LCD instantaneous flow + totalizer on electrical build |
| Process connection | Metric thread M20×1.5 / M36×1.5 standard for DN15 / DN20; flange PN16 or PN25 by order |
| Form factor | Integrated (1) / remote split (2, electrical builds only) |
| Hazardous area | Passive (no Ex cert needed) on local pointer build; Ex d IIC T6 Gb on electrical build |
| Ingress protection | IP65 (electrical build); IP67 / IP68 by special order |
| Power supply | None (mechanical pointer) or 24 VDC / 220 VAC (electrical) |
| Output signals | None (mechanical) / 4-20 mA / HART / RS-485 Modbus / Hi-Lo alarm contacts |
| Approvals | ISO 9001 calibration certificate with each meter; Ex d IIC T6 Gb on electrical Ex build |
| Lead time | 5–7 business days from factory |
Installation
Five install rules. Pre-condition: medium must be clean single-phase within the wetted-parts material window; particulate-laden or 2-phase service is outside the operating range.
- Vertical mount, flow bottom-up; mandatory The metal-tube rotameter relies on gravity to set the float reference; the float rises against gravity in proportion to upward flow. Plumb the body within ±2° of true vertical (spirit-level on the flange). Tilt biases the reading. Horizontal service needs a factory horizontal-build version with a different float; specify at order entry, not at install.
- Upstream filter on clean-service line Float-to-tapered-tube clearance is tight on DN15 / DN20. Particulate larger than ~100 µm can jam the float or score the bore. Install a Y-strainer or basket filter sized for the line, especially on liquid additive or sample-line service where carry-over from upstream tanks is possible.
- Avoid pulsation source upstream VA meters integrate over float oscillation; pulsation causes pointer jitter or biased reading. Mount downstream of any pressure regulator or pump pulsation dampener. For reciprocating dosing-pump service, add an accumulator upstream of the meter to smooth the inlet.
- Light straight-pipe requirement Allow ~5D upstream and ~250 mm (~2D) downstream of any disturbance; elbow, valve, reducer, tee. Lighter than turbine, but still a real requirement. Do not connect a control valve or elbow directly to the meter inlet flange or thread.
- Magnetic interference and temperature shock External magnetic sources above ~3 mT at the indicator bias the magnetic-coupling readout; do not mount near large motors, solenoid valves, or magnetic strainer banks. Avoid rapid thermal cycling (> 50 °C step in < 1 min); thermal stress can warp the magnetic coupling. Allow gradual warm-up on steam or hot-oil service.
Frequently asked questions
Need help sizing a low-flow lab or dosing build?
Send line size, medium, flow range, indicator preference. Reply in 1 day; lead time 5–7 business days.