Precision Handheld Ultrasonic Flow Meter (±0.5 %, DN25–DN5000)
Field Kit for Commissioning Walks, Audit Spot-Checks, and Billing Reconciliation
Transit-time external clamp-on, accuracy spec ±0.5 % at 0.5 to 5 m/s (relaxes to ±1.0 % outside that spec)
24 V DC rechargeable, 5 W. IP54 main unit, IP68 sensor head
Modbus RTU, 4-20 mA, OCT (open-collector transistor) pulse, relay
- P6 grade; covers plant lines up to DN2000, 2.8-inch IPS LCD, 0.20 % repeatability.
- P8 grade; covers trunk lines up to DN5000, 3.5-inch IPS LCD, 0.15 % repeatability (tightest in the family).
- For routine service-call check at ±1.0 %, go to the standard portable family.

The Precision Handheld Ultrasonic Flow Meter is a transit-time field kit for commissioning verification, audit spot-check, and ISO-traceable cross-check of fixed primary meters. External clamp-on install; no pipe cut, no shutdown. ±0.5 % of reading at 0.5 to 5 m/s. P6 grade covers DN25-DN2000 with 2.8-inch IPS LCD and 0.20 % repeatability; P8 grade covers DN25-DN5000 with 3.5-inch IPS LCD and 0.15 % repeatability. Modbus RTU, 4-20 mA, OCT pulse. 24 V DC rechargeable. 5–7 business days.
Benefits
- ±0.5 % of rate, ISO 9001 cal cert in the kit. Tight enough to cite on commissioning handover, ISO 50001 audit, district-heat warranty walks, and disputed-billing reconciliation. Tighter than the standard portable’s ±1.0 %.
- P6 or P8 grade; pick by pipe range and repeatability need. P6: up to DN2000, 0.20 % repeatability. P8: up to DN5000, 0.15 % repeatability with 3.5-inch IPS for outdoor field-service readability.
- External clamp-on install, no pipe cut, no shutdown. Coupling gel + stainless straps; ~30 to 60 minutes per pipe. Same physics as fixed clamp-on; portable form factor.
- IP68 sensor head, IP54 main unit. Sensor head tolerates submersion at the install point; main unit tolerates field-service rain and splash without ingress.
- 24 V DC rechargeable with universal 90-260 V AC dock. Full-day service on battery; overnight recharge. Modbus RTU + 4-20 mA + OCT pulse. Carry case included. 5–7 business days.
Configure your build.
Build a model code from grade, output, accessory, and quantity.
Your configuration
Applications
HVAC commissioning validation
Plant handover walks on chiller primary and condenser. Cite ±0.5 % spot-check to clear inline meters at install. Cal cert in the kit.
District-heat warranty audit
Substation primary cross-check during warranty period. P8 grade suits DN above 2000 mains and reports the 0.15 % repeatability on the audit form.
ISO 50001 EnMS (Energy Management System) cross-check
Quarterly audit walks; primary-meter drift verification. Defensible to ±0.5 % against challenge in EnMS review.
Disputed billing reconciliation
Utility billing dispute, primary meter suspected to drift. Field cross-check produces evidence customers accept.
Full specifications.
| Measurement principle | Transit-time ultrasonic; external clamp-on transducer pair; firmware times the upstream and downstream transit difference |
|---|---|
| Grade | P6 (DN25-DN2000, 2.8-inch IPS LCD 320 x 240, 0.20 % repeatability) OR P8 (DN25-DN5000, 3.5-inch IPS LCD 480 x 320, 0.15 % repeatability) |
| Pipe diameter range | DN25 to DN5000 on P8 grade; DN25 to DN2000 on P6 grade |
| Nominal diameter | Clamp-on external mount; the kit itself has no process port; the host pipe DN is the measured diameter |
| Channels | 1 acoustic path on the handheld kit (single-path transit-time) |
| Install style | External clamp-on field kit with stainless straps; coupling gel between transducer face and host pipe outer surface |
| Optional BTU pair | Not included on the precision handheld; go to the standard portable (PT100 pair on-board) or the inline BTU meter when energy output is part of the audit citation |
| Power | 24 V DC, 5 W; rechargeable Li-ion battery with universal mains charging dock |
| Accuracy | ±0.5 % of reading at velocity 0.5 to 5 m/s on both grades; relaxes to ±1.0 % below 0.5 m/s and above 5 m/s |
| Linearity | ±0.5 % at velocity 0.5 to 5 m/s |
| Repeatability | 0.20 % of span (P6); 0.15 % of span (P8) |
| Velocity range | 0.1 to 12 m/s operating, bi-directional with reverse-flow detection |
| Response time | 500 ms |
| Display | 2.8-inch IPS LCD 320 x 240 (P6) OR 3.5-inch IPS LCD 480 x 320 (P8); 4-direction rotation; unit selection m³, L, GAL combined with day, hour, min, sec time units |
| Output; analog | 4-20 mA, max load 600 ohm |
| Output; digital | RS-485 Modbus RTU |
| Output; other | OCT pulse + high / low alarm; relay 30 V DC at 1 A |
| Power supply | 24 V DC, 5 W; rechargeable Li-ion battery with universal 90-260 V AC mains charging dock; bench supply for desk verification |
| Power consumption | 5 W typical at 4-20 mA output mode |
| Medium | Single-phase liquid, 100 % pipe-fill; water, aqueous solution, chemical reagent; suspended solids 4 % or less |
| Medium temperature | −40 to +60 deg C main unit; −40 to +80 deg C sensor head |
| Ambient temperature | −40 to +60 deg C main unit operating |
| Ingress protection | IP54 main unit; IP68 sensor head (submersible at install point) |
| Hazardous area | Not declared on this model; route Ex-classified service to the fixed clamp-on or insertion counterpart with CNEX Ex d IIC certificate |
| Process connection | External clamp-on; no process port on the kit; mounts on the host pipe outer surface with stainless straps |
| BTU calorimetry | Not included on the precision handheld; for BTU field record, go to the standard portable (PT100 pair) or to a fixed BTU meter |
| Compliance ships standard | ISO 9001 calibration certificate with each kit (serial number + tested velocity range + traceability note); EMC 2014/30/EU; carry case included |
| Lead time | 5–7 business days from factory |
Installation rules.
- 10D upstream, 5D downstream straight-pipe rule. Pick the longest available straight run at the audit point.
- Prep the host-pipe outer surface. Wipe paint, rust, and insulation off the transducer footprint. Apply coupling gel; secure with stainless straps at the specified spacing.
- Verify 100 % pipe-fill. Air pockets falsify the signal; closed loops need a vent on the loop high point upstream of the audit point.
- Check battery state and cal-cert serial at the start of each walk. Log the cal-cert serial on the audit record; kit ships with traceability back to ISO 9001 reference master.
- Recharge overnight on the universal mains dock. 90 to 260 V AC, no rewiring at international sites.